Absorbed dose in the operator’s brain in interventional radiology practices: evaluation through KAP value conversion factors (CROSBI ID 281390)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

(EURADOS Working Group) Ferrari, Paolo ; Jovanović, Zoran ; Bakhanova, Elena ; Becker, Frank ; Krstić, Dragana ; Jansen, Jan ; Principi, Sara ; Teles, Pedro ; Clairand, Isabelle ; Knežević, Željka Absorbed dose in the operator’s brain in interventional radiology practices: evaluation through KAP value conversion factors // Physica medica, 76 (2020), 177-181. doi: 10.1016/j.ejmp.2020.07.011

Podaci o odgovornosti

Ferrari, Paolo ; Jovanović, Zoran ; Bakhanova, Elena ; Becker, Frank ; Krstić, Dragana ; Jansen, Jan ; Principi, Sara ; Teles, Pedro ; Clairand, Isabelle ; Knežević, Željka

EURADOS Working Group

engleski

Absorbed dose in the operator’s brain in interventional radiology practices: evaluation through KAP value conversion factors

In order to address the recent concerns over a possible increasing in brain tumour mortality among interventional radiologists and cardiologist, this work evaluated the exposure conditions of the operator’s brain during interventional procedures using Monte Carlo simulations with anthropomorphic phantoms. The absorbed doses in several predefined segments of the operator’s brain were estimated in a typical interventional radiology irradiation scenario. The doses were normalized to the KAP values simulated for ten X-ray beam qualities and four projections (PA, RAO 25°, LAO 25° and CRA 25°). For the interventional radiology scenario, because of the position of the operator, no difference was found in the exposure between the left and right regions of the brain for the first operator. However, for the second operator standing at a farer distance from the tube, the exposure of the left part of the brain is up to two times higher than that of the right part. The results are in agreement with dose measurements reported in the literature. The conversion factors, obtained as the absorbed dose per KAP, can be used to obtain a first estimate of the exposure of the brain of the operators during interventional procedures.

Brain exposure ; Absorbed dose ; Monte Carlo simulation ; Medical staff dosimetry ; Interventional radiology practice

WG12 Dosimetry in medical imaging

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

76

2020.

177-181

objavljeno

1120-1797

1724-191X

10.1016/j.ejmp.2020.07.011

Povezanost rada

Fizika, Interdisciplinarne prirodne znanosti, Kliničke medicinske znanosti