Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

Cochlear implant electrode array extrusion – is reimplantation feasible? – case report (CROSBI ID 274899)

Prilog u časopisu | prikaz, osvrt, kritika | domaća recenzija

Velepič, Marko ; Manestar, Dubravko ; Valković Zujić, Petra ; Malvić, Goran ; Zubović Velepič, Sanja Cochlear implant electrode array extrusion – is reimplantation feasible? – case report // Medicina Fluminensis, 55 (2019), 1; 99-104

Podaci o odgovornosti

Velepič, Marko ; Manestar, Dubravko ; Valković Zujić, Petra ; Malvić, Goran ; Zubović Velepič, Sanja

engleski

Cochlear implant electrode array extrusion – is reimplantation feasible? – case report

Aim: To present two cases of late complications of cochlear implantation. Case report: In both cases extracochlear extrusion of the implant electrode occurred. More than ten years after successful cochlear implantation, the hearing performance gradually worsened to the point of no intelligibility in both patients. Otomicroscopy revealed chronic otitis media (COM) with an electrode in the external ear canal in implanted ears of both patients. Extracochlear electrode extrusion was noticed on CT scans in both cases. In the first case the device was explanted, the electrode was left in the cochlea and tympanoplasty was performed. Regardless the normal status of the middle ear, extracochlear extrusion occurred again. Conclusion: After extracochlear electrode extrusion, the reimplantation is likely to be unsuccessful.

chronic otitis media ; cochlear implant ; extracochlear electrode extrusion

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

55 (1)

2019.

99-104

objavljeno

1847-6864

Povezanost rada

Kliničke medicinske znanosti

Indeksiranost