Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

Verification study of free light chains assays on reagent-optimized analysers (CROSBI ID 272117)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Šegulja, Dragana ; Matišić, Danica ; Barišić, Karmela ; Rogić, Dunja Verification study of free light chains assays on reagent-optimized analysers // Biochemia medica, 29 (2019), 3; 579-586. doi: 10.11613/bm.2019.030709

Podaci o odgovornosti

Šegulja, Dragana ; Matišić, Danica ; Barišić, Karmela ; Rogić, Dunja

engleski

Verification study of free light chains assays on reagent-optimized analysers

Our aim was to compare analytical specifications of two assays (monoclonal vs. polyclonal) for free light chains (FLCs) quantification optimized for two different analytical platforms, nephelometer ProSpec (Siemens, Erlangen, Germany) and turbidimetric analyser Optilite (The Binding Site, Birmingham, UK). The evaluation included verification of the precision, repeatability and reproducibility, estimation of accuracy and method comparison study with 37 serum samples of haematological patients. Kappa and lambda FLC were measured in each sample by both methods and kappa/lambda ratio was calculated. Results show satisfactory precision of both methods with coefficients of variation for ProSpec of CVwr = 2.20% and CVbr = 3.44%, and for Optilite CVwr = 2.82% and CVbr = 4.15%. Estimated bias for FLC lambda was higher on the ProSpec analyser, but bias for FLC kappa was higher on the Optilite analyser. Correlation coefficients were 0.98 ; P < 0.001 for FLC kappa and 0.97 ; P < 0.001 for FLC lambda. Considering normal/pathological FLC ratio moderate agreement within assays was detected (κ = 0.621). When the results were categorized according to criteria for progressive disease, 4/37 (0.10) cases were differently classified. Lambda FLC values by Optilite in three samples with monoclonal FLC lambda were more than twelve times higher than by ProSpec. A 25% difference in FLC ratio was detected in 16/37 (0.43) and 50% difference in 13/37 (0.35) patients. All manufacturers’ precision claims could not be achieved in the verification study. The comparison of results to biological variations data showed that coefficients of variations are acceptable for both assays. The assays should not be used interchangeably in haematological patients.

M components ; serum free light chains ; immunoassay ; gammopathy ; monoclonal

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

29 (3)

2019.

579-586

objavljeno

1330-0962

10.11613/bm.2019.030709

Povezanost rada

Biotehnologija u biomedicini (prirodno područje, biomedicina i zdravstvo, biotehničko područje), Kliničke medicinske znanosti

Poveznice
Indeksiranost