Scintillator Pixel Detectors for Measurement of Compton Scattering (CROSBI ID 266893)
Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Makek, Mihael ; Bosnar, Damir ; Pavelić, Luka
engleski
Scintillator Pixel Detectors for Measurement of Compton Scattering
The Compton scattering of gamma rays is commonly detected using two detector layers, the first for detection of the recoil electron and the second for the scattered gamma. We have assembled detector modules consisting of scintillation pixels, which are able to detect and reconstruct the Compton scattering of gammas with only one readout layer. This substantially reduces the number of electronic channels and opens the possibility to construct cost-efficient Compton scattering detectors for various applications such as medical imaging, environment monitoring, or fundamental research. A module consists of a 4 × 4 matrix of lutetium fine silicate scintillators and is read out by a matching silicon photomultiplier array. Two modules have been tested with a 22Na source in coincidence mode, and the performance in the detection of 511 keV gamma Compton scattering has been evaluated. The results show that Compton events can be clearly distinguished with a mean energy resolution of 12.2% ± 0.7% in a module and a coincidence time resolution of 0.56±0.02 ns between the two modules.
Compton camera ; positron emission tomography ; Compton scattering ; scintillation detector ; silicon photomultiplier ; medical applications